Automated External Defibrillator KED-A100 is a biphasic portable semi-automated public access defibrillation device. System incorporates advanced cardiac life-support features which include step-by-step voice instructions from pad placement to shock delivery with clear visual real-time guidance indicators as well as electrode position analysis for safe and quick treatment aid during cardiac emergency suitable to use for on-the-spot responders with or without training.
| Waveform | Biphasic |
| Energy Sequence | 150 J, 200 J |
| Analysis Time | 9 seconds |
| Audio Indications | 20 distinct voice instructions |
| Charge Time | 8 seconds for 150 J 10 seconds for 200 J |
| Capacity | 100 discharges (200 J) 120 discharges (150 J) |
| Rhythm Analysis Time | Less than 30 secs for new battery Less than 35 secs after 6 shock discharge Less than 25 secs at initial switch-on |
| Visual Indications | LED flash light indicator |
| Control Keys | Shock and On/Off keys |
| Energy Accuracy | ± 15 % tolerance (25 Ω to 175 Ω) |
| Impedance Range | 20 Ω to 200 Ω |
| Battery | Non-rechargeable (12V DC) |
| Operating Temperature | 0 °C to 40 °C |
| Operating Humidity | 30 % to 95 % |
| Dimensions | 303 x 216 x 89 mm |
| Weight | 2 Kgs |
- Simple 3-step defibrillation process
-
Dual switch-press control
-
Max. discharge energy up to 200 J
-
Up to 20 distinct voice prompts
-
Lock-out protection facility
-
LED visual indicators